ECG | |
---|---|
Normal Sinus Rhythm | |
Rates | 30, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM |
Rate Accuracy | ±1BPM or 1% |
Amplitude (lead II) | 0.05 to 0.45mV (0.05mV steps)[br]0.5 to 5.5mV (0.5mV steps) |
Amplitude Accuracy | ± 2% |
ST Segments | 18 selectable values, 8 elevated and 8 depressed -0.8 to +0.8mV in 0.1mV steps and +0.05mV and -0.05mV on Lead II |
Neonatal Mode | ECG R wave width is reduced to 40ms |
Pacemaker | |
---|---|
Simulated Rhythms | Asynchronous at 75 BPM[br]Demand with frequent sinus beat[br]Demand with occasional sinus beat[br]A-V sequential[br]Non-capture[br]Non-function |
Pulse Amplitude | 1.0, 2.0, 5.0, 10.0mV |
Accuracy | ± 5% |
Width | 0.1, 0.2, 0.5, 1.0, 2.0ms |
R-Wave Detector Test | 30, 60, 80, 120, 200, 250 BPM |
Haver-Triangle Width | 8, 10, 12, 20, 40 ,60, 80, 100, 120, 140, 160, 180, 200ms |
Performance Amplitude | 0.5 to 5.0mV in 0.5mV steps |
Fetal Simulation | |
---|---|
Maternal Heart Rate | 80 BPM (Fixed) |
Fetal Heart Rate (Selectable) | 60, 90, 120, 140, 150, 210 and 240 BPM Fetal Heart Rate (IUP) 140 BPM at beginning, then varying with pressure |
pressure Intrauterine-Pressure Waveforms | Early deceleration, late deceleration and uniform acceleration |
Simulation Period | Manual, or 2, 3, or 5 minutes |
Temperature | |
---|---|
Simulation | YSI 400 / 700A / 700B Static |
Temperature Unit | °C or °F, user selectable |
Range | 0.0, 24.0, 37.0, and 40.0°C[br]32.0, 75.2, 98.6, 104.0°F |
Accuracy | ± 0.1°C / °F |
Connector | Mini DIN style |
Mains Power/Battery Info | |
---|---|
Mains Power/Battery Info | 3.7V 3900mAh 14.4WH[br]5V 1A USB micro-B power supply[br]100-240V ~ 50/60Hz 0.18A max. |
Charge Time (new battery) | Up to 3-4 hours |
Battery Life | Up to 6 hours, ~4 hours |
Weight | 0.70kg / 1.5lbs |
Dimensions | 180 x 150 x 55mm, 7.1 x 5.9 x 2.2" |
Simulation Type | |
---|---|
Lead I | 70% |
Lead II | 100% |
Lead III | 30% |
Lead V1 | 24% |
Lead V2 | 48% |
Lead V3 | 100% |
Lead V4 | 120% |
Lead V5 | 112% |
Lead V6 | 80% |
ECG Performance Tests | |
---|---|
Square Waves | 2Hz, 0.125Hz |
Triangle Wave | 2Hz |
Pulse | 60bpm or 240bpm |
Sine Waves | 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100Hz |
Performance Amplitude | 0.5 to 5.0mV in 0.5mV steps |
Arrhythmias | |
---|---|
Ventricular | Asystole[br]Bigeminy[br]Trigeminy[br]Ventricular fibrillation (coarse)[br]Ventricular fibrillation (fine)[br]Ventricular tachycardia |
Supraventricular | Atrial fibrillation (coarse)[br]Atrial fibrillation (fine)[br]Atrial flutter[br]Atrial tachycardia[br]Missing beat[br]Nodal rhythm[br]Paroxysmal atrial tachycardia[br]Sinus arrhythmia[br]Supraventricular tachycardia |
Atrial Conduction | First degree AV block[br]Left bundle branch block[br]Right bundle branch block[br]Second degree AV block - Mobitz I[br]Second degree AV block - Mobitz II[br]Third degree AV block |
Premature | Premature atrial contraction[br] Premature nodal contraction[br] Premature left ventricle contraction[br] Premature left ventricle contraction - early[br] Premature left ventricle contraction - R on T[br] Premature right ventricle contraction[br] Premature right ventricle contraction - early[br] Premature right ventricle contraction - R on T[br] Premature ventricular contraction - 6 / min[br] Premature ventricular contraction - 12 / min[br] Premature ventricular contraction - 24 / min[br] Premature ventricular contraction - frequent multifocal[br] |
Invasive Blood Pressure | |
---|---|
Channels | 2 channels |
Static Pressure | -10, -5, 0, 20, 40, 50, 60, 80,100, 150,160, 200, 240, 320, 400mmHg |
Dynamic Simulation | Arterial [ART] 120/80[br] Radial Artery [RA] 120/80[br] Left Ventricle [LV] 120/00[br] Right Ventricle [RV] 25/00[br] Right Atrium (central venous) [CVP] 15/10[br] Pulmonary Artery [PA] 25/10[br] Pulmonary Artery Wedge [PAW] 10/2[br] Left Atrium [LA] 14/4 |
Accuracy | ± 1mmHg |
Excitation voltage | 2 to 16V |
Impedance | 350Ω Nominal |
Simulated sensitivity | 5μV/V/mmHg or 40μV/V/mmHg (user selectable) |
Connector | mini DIN style |
Respiration | |
---|---|
Rates | Selectable values of 0, 5, 10, 15, 30, 40, 60, 80, 120, 180 brpm |
Lead Selection | Lead I and II, user selectable |
Base Resistances | 500, 1000, 1500 & 2000Ω |
Accuracy | ±10% |
Resistance Variations | 0.2, 0.5, 1.0, 3.0Ω |
Apnoea Simulation | Manual ON / OFF |
Environmental Specifications | |
---|---|
Operating Conditions | 10 - 40°C (50 - 104°F) 0-90% RH - NC |
Storage Environment | -15 - 60°C (5 - 140°F) 0-90% RH - NC |
Environmental Protection | IP40 |