Citrex H3
– Test and calibrate ventilators
– Flexible trigger settings and an optional oxygen sensor
– Built-in pressure and gas flow sensors
– Measure and display 16 different ventilation parameters
– Test and calibrate ventilators
– Flexible trigger settings and an optional oxygen sensor
– Built-in pressure and gas flow sensors
– Measure and display 16 different ventilation parameters
Manual
Manuals
Datasheets
Pressure | |
---|---|
Flow channel | -50–150 mbar ±0.75 %* or ±0.1 mbar** |
Atmospheric pressure | 500–1150 mbar ±1 %* or ±5 mbar** |
Flow and Pressure Measurements | |
---|---|
Measuring direction | Bidirectional |
Flow | ±300 sL/min*** ±1.9 %* or ±0.1 sL/min (for 10..40°C)** |
Temperature compensated | Automatic |
Pressure compensated | Automatic |
Humidity compensated | Manual |
Other measurements | |
---|---|
Oxygen (Option) (pressure comp. ≤ 150 mbar) | 0–100 % ±1 % O2** |
Gas temperature | 0–50°C ±1.75 %* or ±0.5 °C** |
Gas types | Air, Air/O2, O2, N2O, N2O/O2 |
Gas standards | ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD |
Units | |
---|---|
Flow | L/min, L/s, cfm |
Pressure | inH2O, bar, mbar, cmH2O, mmHg |
General information | |
---|---|
Display | 1.7” with touch control elements (color), Realtime curves |
Data storage | microSD Card |
Power | 11.4 x 7 x 6 cm |
Weight | 0.38 kg |
Battery | 4 hours |
Approvals | CE, BC (Energy Efficiency for Battery Charging Systems), CSA (North America), IEC 61010-1:2010, IEC 61326-2:2012 |
Ventilation parameter | |
---|---|
Breath rate | 1–1000 bpm ±1 bpm* or ±2.5 %** |
Time | 0.05–60 s ±0.02 s |
Ratio | 1:300–300:1 ±2.5 %* |
Tidal Volume | ±10 L ±2 %* or ±0.20 mL (>6 sL/min)** |
Minute Volume | 0–300 sL/min ±2.5 %* |
Peak flow | ± 300 sL/min ±2.0 %* or ±0.1 sL/min** |
Pressure | 0–150 mbar ±0.75 %* or ±0.1 mbar** |
The greater tolerance is valid: * Tolerance related to the measured value, ** Absolute tolerance, *** The unit sL/min is based on ambient conditions of 0 °C and 1013.25 mbar (DIN 1343). |
Register your product today to activate your warranty and to stay up to date with the latest product updates, news and support information.
If you wish to calibrate your Citrex H3, please locate the nearest service center or contact us directly using the link below.
Sign up to our mailing list today to stay up to date with the latest industry news and information from Seaward.