Pressure | |
---|---|
High | -1–10 bar ±1 %* or ±7 mbar** |
Differential | ±200 mbar ±0.75 %* or ±0.1 mbar** |
Flow channel | -50–150 mbar ±0.75 %* or ±0.1 mbar** |
Atmospheric pressure | 500–1150 mbar ±1 %* or ±5 mbar** |
Flow and Pressure Measurements | |
---|---|
Flow | ±300 sL/min ±1.9 %* or ±0.1 L/min (for 10..40°C)** |
Measuring bidirectional | yes |
Temperature compensated | automatic |
Pressure compensated | automatic |
Humidity compensated | manually |
Other measurements | |
---|---|
Oxygen (pressure comp. ≤ 150 mbar) | 0–100 % ±1 % O2** |
Gas temperature | 0–50°C ±1.75 %* or ±0.5 °C** |
Gas types | Air, Air/O2, O2, N2O, N2O/O2, CO2, N2, He/O2 |
Gas standards | ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD, BTPD-A, 0/1013, 20/981, 15/1013, 25/991, 20/1013, NTPD, NTPS |
Units | |
---|---|
Flow | L/min, L/s, cfm, mL/min, mL/s |
Pressure | bar, mbar, cmH2O, inH2O, Torr, inHg, hPa, kPa, mmHg, PSI |
General information | |
---|---|
Realtime curves | yes |
Display | 4.3” Multi-Touch (color) |
Interface | RS-232, USB, Ethernet, CAN, Analog Out, TTL, WLAN |
Data storage | Internal and microSD Card |
Power | 100–240 VAC, 50–60 Hz |
Dimension (w x d x h) | 11.4 x 7 x 7,3 cm |
Weight | 0.52 kg |
Battery | 5 hours |
Approvals | CE, BC (Energy Efficiency for Battery Charging Systems), CSA (North America), IEC 61010-1:2010, IEC 61326- 2:2012 |
Ventilation parameter | |
---|---|
Breath rate | 1–1000 bpm ±1 bpm* or ±2.5 %** |
Time | 0.05–60 s ±0.02 s |
Ratio | 1:300–300:1 ±2.5 %* |
Ti/Tcyc | 0–100 % ±5 %* |
Breath volume | ±2 %* or ±0.20 mL (>6 sL/min)** |
Tidal volume | ±10 L ±2 %* or ±0.20 mL (>6 sL/min)** |
Minute volume | 0–300 sL/min ±2.5 %* |
Peak flow | ± 300 sL/min ±1.9 %* or ±0.1 sL/min** |
Pressure | 0–150 mbar ±0.75 %* or ±0.1 mbar** |
Compliance | 0–1000 mL/mbar ±3 %* or ±1 mL/mbar** |