Insulation Resistance | |
---|---|
Measurement | EUT to earth/ground, EUT to AP, AP to earth/ground |
Voltage | 50 / 100 / 250 & 500V DC @1mA |
Range (low range) @ 50V DC | 0.01Mohms - 10 Mohms |
Range (low range) @ Above 50V DC | 0.01Mohms - 20Mohms |
Accuracy (low range) | ± 5% of reading +2 counts |
Range (high range) @ 250V DC | 20Mohms – 50Mohms |
Range (high range) @ 500V DC | 20Mohms – 100Mohms |
Accuracy (high range) | ±10% +2 counts |
Resolution | 0.01Mohms |
Differential Leakage Measurement | |
---|---|
Measuring Range | 75μA to 9999μA |
Accuracy | ±5% of reading + 5 counts |
Measurement / display resolution | 1μA |
Measurement Type | True RMS |
Measuring Device | Frequency response characteristics to IEC 60601-1 |
Power Measurement | |
---|---|
Range | 0.1kVA – 4kVA |
Accuracy | ±10% + 2 counts |
IEC Mains Lead Test | |
---|---|
Test Duration | 2s |
Test | Continuity of all conductors, Insulation and Polarity |
Standard Accessories (supplied with 288+) | |
---|---|
Calibration certificate | |
Earth/ground bond test probe with clip | |
Earth/ground bond clip lead | |
Patient Applied part module | |
10 Applied part adaptors | |
Detachable 2 meter mains cable | |
Bluetooth USB dongle | |
Electronic instruction manual | |
Quick start guide | |
Application software |
Earth Continuity | |
---|---|
Method | 2 wire technique, using 'zero' lead function. |
Test Current | >+200mA, -200mA DC into 2 ohms |
Max Test Voltage | 4-24V RMS o/c |
Measuring Range (low range) | 0.001 – 0.999 ohms @ 0.001 ohms resolution |
Measuring Range (mid range) | 1.00 – 9.99 ohms /@ 0.01 ohms resolution |
Measuring Range (high range) | 10.0 – 19.9 ohms @ 0.1 ohms resolution |
Accuracy | ± 3% of reading + 10 m ohms |
Direct Leakage Measurement | |
---|---|
Measuring Range | 4μA to 9999μA |
Accuracy | ± 5% or reading +2 counts |
Mains on A.P. voltage | F-type only @ 110% of mains |
Measuring Device | As per IEC 60601-1 requirements |
Measurement Type | Separate AC & DC for Patient (Auxiliary) Leakage to IEC 60601, True RMS for all remaining Leakage tests |
Alternative Leakage Measurements | |
---|---|
Test Voltage | 250V at mains frequency |
Test Current | 3.5 mA current limited |
Measurement Range | 4μA to 9999μA |
Measurement Resolution | 1μA |
Measurement Accuracy | ±5% of reading + 2 counts |
Measurement Type | True RMS |
Measuring Device | As per IEC 60601-1 |
Mains Outlet Test | |
---|---|
Input voltage range | 0-300V AC |
Max current | 16A |
Measures | L – E, N – E |
Accuracy | ± 5% of reading + 2 counts |
General Specification | |
---|---|
Mains Power | 230V AC ±10%, 50-60Hz +/- 1Hz, 120V AC ±10%, 60Hz +/- 1Hz (USA model) |
Battery | 6 x 1.5V AA |
Weights | 1.6 kg / 3.5 lbs including batteries |
Size | (L x W x D) 270 x 110 x 75 mm / 10.5 x 4 x 3” |
Operating Conditions | 0'C - 40'C, 32'F - 104'F, 0-90% RH - NC |
Storage Environment | -15'C - +60'C / 5'F - 140'F |
Environmental Protection | IP 40 |