Rigel UNI-SIM

Vital Signs Simulator (NIBP / SPO2 and ECG)

  • Accurate 6 in 1 vital signs simulation
  • Synchronised heart rate
  • Hand-held and Battery powered
  • On-board memory and keyboard
  • Manual and automatic operation
  • Onboard data entry and storage
  • Custom test protocols
  • Download / Upload software
Rigel UNI-SIM Vital Signs Simulator
Click to Request a Demonstration / Quote

Rigel UNI-SIM Vital Signs Simulator

The Rigel UNI-SiM Vital Signs Simulator is the world’s first battery powered 6 in 1 vital signs simulator able to undertake six synchronised vital signs parameters tests.

Enabling quick, easy and accurate simulations of NiBP, SpO2, ECG, temperature, IBP and respiration functionality, the Rigel Uni-Sim Vital Signs Simulator is one of the most valuable tools around today.

Utilising new and advanced technologies in both SPO2 ad NIBP simulation, reduces errors and minimises the test time. Manufacturer specific simulation curves can be created and uploaded to further aid the accuracy.

Bluetooth connectivity to a range of accessories and Med-eBase, comprehensive database software, ensures fast and easy download of test results, managing your asset database, creation of test sequences and producing test certificates.

The Rigel PULS-R SpO2 Finger Simulator can also be used with the UNI-SiM vital signs simulator

Rigel Uni-Sim with PULS-R

Click here to view all optional accessories available for the Rigel SiM range

 

 

Free Vital Signs Booklet

We have created an in-depth guide to measuring and simulating vital signs.
 


Download now

Rigel UNI-SIM Vital Signs Simulator

Technical Specification:

Non-invasive Blood Pressure
Simulation Waveform Oscillometric
Integrated Pump 0 to 350 mmHg user configurable
Leak Test User configurable between 0-350mmHg
Chronometer Configurable upto 999 secs
Digital Manometer 0 - 410 mmHg
Pressure Accuracy +/- 0.5% FS
Pressure Units mmHg, inHg, kg/cm2, cmH2O, mBar, PSI, inH2O and kPa
Oxygen Saturation
Range 50 to 100%
Accuracy ± 1 bpm
Chronometer Test option to test response time SpO2 monitor
ECG Arrhythmia Simulator
ECG 5 lead simulation including high level output
Wave Forms Normal Sinus Rhythm (NSR), ST Elevation, ST Depression, Myocardial Infarction, Tall T
Rate 20 – 300 BPM typed value
Performance Waveforms Sine, Square, Triangle, and Pulse
Pacer Waveforms Synchronous Atrial, Asynchronous Atrial
Pacer Only Pacer Pulse rate 60 BPM, Ventricular Pacer rate 70 BPM, Atrial & Ventricular Pacer Heart rate 70 BPM, R Wave Detection
Arrhythmia Waveforms
Atrial Sinus Arrhythmia (SA), Missing beat, Atrial Flutter (AFLT), Atrial Fibrillation (AFB), Paroxysmal Atrial Tachycardia (PAT), Junctional Premature Contraction
Atrial Conduction First Degree AV Block, Second Degree AV Block - Mobitz I, Second Degree AV Block - Mobitz II, Third Degree AV Block, Right Bundle Branch Block (RBB), Left Bundle Branch Block (LBB), Left Anterior Hemiblock
Ventricular Premature Ventricular Contraction - Intermittent Premature Ventricular Contraction – Continuous, Bigeminy, Trigeminy, Ventricular Flutter (VFLT), Ventricular Fibrillation (VFB), Ventricular Tachycardia (VTC), Right Focal PVC
Temperature
Simulation YSI 400 / 700 Static
Range Preset at 25, 33, 37and 41°C
Respiration
Rates 5, 10, 15, 30, 60, 120, 180 Breaths per Minute
Base Resistances 250, 500, 750, 1000 ohms
Apnoea Simulation 0 – 60 seconds duration 0 – 300 seconds interval
Invasive Blood Pressure
Channels 2 channels
Static 0 to 300mmHg. Typed values
Dynamic 0 - 300mmHg for systolic & diastolic
General Specification
Operation Battery cell, insitu charge
Battery Charger 110-230VAC, 50/60 Hz
Battery Life 4-24 hours testing depending on simulations
Memory Capacity Approximately 10,000 records
Communication Via Blue Tooth
Display Monochorme, ¼ VGA full graphics
Keypad Alpha-numeric
Weight <1.5 kg
Size (L x W x D) 270 x 110 x 75 mm / 10.5 x 4 x 3”
Operating Conditions 10-30°C, 0-90% RH - NC
Storage Environment -15°C - +60°CC
Environmental Protection IP 40
Part Number
370A930  

Bluetooth Barcode Scanner

The Bluetooth barcode scanner from Rigel Medical is a must have accessory for those who require fast and accurate data entry into the Rigel product offering. View Product

Bluetooth Results Printer

The battery powered results printer is one of the most compact printers on the market. Able to print results directly from the Rigel testers onto a 50mm wide paper roll. View Product

Field Service Kit

Housed in a hard wearing, impact and water resistant Pelican carry case, the Rigel field service kit is a handy and more specialised carrying solution. View Product

Med-eBase

This asset management program facilitates the production of a comprehensive series of advanced results and data reports. Download a free trial today. View Product

Med-eCare

Rigel’s Med-eCare Plan is at your service for the entire Rigel Medical range. It covers all typical and unexpected costs associated with calibration and maintenance of your test equipment for the duration of the plan. View Product

Med-eKit

Rigel Medical has introduced a new trolley case-based biomed test kit used for the rapid testing of electrical safety and operation of medical devices. View Product

Rigel Med-ePack

The Rigel Med-ePack solution package offers a complete test set that could include an electrical safety analyser, vital signs simulator, ventilator tester and more. View Product

Sling-style Carry Case

This new case is a standard accessory for the Rigel 288 and UNI-SIM. It can be configured to hold a number of different items of test equipment and accessories. View Product

Test n Tag Elite 2 Printer

Rigel's Test n Tag Elite 2 printer is a unique accessory that provides you with print-outs of test results and pass/fail barcode labels and has bluetooth connectivity. View Product

Contact Us
       