Calibration Checkbox
To quickly verify the calibration of medical electrical safety analysers, Rigel Medical has introduced the Rigel 601 calibration checkbox.
|
Earth Bond
Test Points 0.05 Ohm *
0.10 Ohm *
0.20 Ohm *
Accuracy 2%, nominal value, factory calibrated
Insulation
Test Points 2M Ohm
20M Ohm
Accuracy 1%, nominal value, factory calibrated
Leakage
PASS value 15% under IEC 60601-1 test limit (See table T1)(Earth / Enclosure / Patient)
FAIL value 15% over IEC 60601-1 test limit (See Table T1) Accuracy 0.5%, factory calibrated
Leakage (F-type Mains on Applied Part)
PASS value 25uA
FAIL value 2.5mA
Accuracy Nomilnal value,
General
Calibration 36 months
Warranty 24 months
Operating Conditions 15-30’C, 0-98% RH Non condensating
Supplied with
Certificate of calibration
Function earth cable
Instruction Manual