.jpg)
The Most Complete Evaluation and Control Portable Tool
The Rigel 355 is a unique hand-held ventilator tester for both adult and paediatric ventilators. The Rigel 355 offers highly accurate measurements of: Respiration Frequency, Tidal volume, Maximum Pressure, Minimum Pressure, Ventilation, Expiratory average flow, Inspiratory average flow and Ratio I /E.
The Rigel 355 has a build-in dynamic memory to store performance graphs for download to your PC. When used with the included RESPI software, the Rigel 355 offers a full stand-alone ventilator test system which provides real time graphs on your PC.
The RESPI software provides additional parameters such as: Maximum Expiratory Flow, Tidal Volume of Inspiratory and Expiratory, Inspiratory Average Pressure, PEEP, Index of Internal and External Resistance, Resistance and Compliance of the Passive lung, Inspiratory and Expiratory Time.
|
Hand-Held |
Paediatric and Adult |
|
Battery Powered |
Large LCD display |
Mecanical and electrical characteristics of Rigel 355
Dimensions : 225 x 115 x 80 mm ( + external pneumotachograph)
Weight : 0.900 Kg
Supply voltage : internal battery (12 hours autonomy)
mains supply with a 230 Volts adapter
Accessories
-Carrying suitcase
-Pneumotachographs (from n° 00000 to n° 5)
-Portable pr inter connected directly on Rigel 355
-Car charger
- Passive lung
Parameters directly calculated and displayed on Rigel 355 :
| Parameters | Symbol | measures |
Precision |
Display Resolution |
| Respiratory Frequency | FR |
- |
- |
0.01 cycle/min |
| Ratio I /E | I /E | 0.01 to 9.99 |
- |
0.01 |
| Tidal volume | Vt | Limited by the volume of the charge |
- |
10 mL |
| Maximum Pressure | Pmax | FS : + /- 75 hPa | 0.01 hPa | |
| Minimum Pressure | Pmin | FS : + /- 75 hPa | + /- 1% | 0.01 hPa |
| Ventilation | L/min | - | + /- 1% | 10 mL/min |
| Expiratory average flow | Fmoy | Depends on the used pneumo | + /- 1% | 10 mL/s |
| Inspiratory average flow | Fmoy | Depends on the used pneumo | + /- 1% | 10 mL/s |
Parameters displayed in Manometer / Flowmeter Mode:
| Parameters | Symbol | measures |
Precision |
Display Resolution |
| Flow (pneumo type 2) | D |
± 2,5 L/s or 150 L/min |
2% | 0.01 L/min |
| Flow (pneumo type 0) | D | ± 0,25 L/s or 15 L/min | 2% | 0.1 mL/min |
| Low Pressure | Pb | ± 60 hPa | 1% | 0.01 hPa |
| High Pressure | Ph | 0 à 5 bar s | 1% | 0.001 bar |
| FiO2 | O2 |
+ /- 100 % |
1% |
0.01 hPa |
Calculated parameters by Rigel Respi software ( Rigel 355 used in Ventest Mode)
| Parameters | Range of Measures | Precision | Display Resolution |
|
Maximum Pressure |
+/- 75 hPa | +/- 1% | 0.01 hPa |
| Maximum expiratory flow Inspiratory |
originally +/- 3 L/s (depends on the used pneumo, see table) |
+/- 5% (for 1 L volume) | 10mL/s |
| Expiratory Average flow Inspiratory |
originally +/- 3 L/s (depends on the used pneumo, see table) |
+/- 5% (for 1 L volume) | 10mL/s |
| Tidal Volume (Insp and Exp) | Depends on the volume of the charge | 10mL | |
| Ratio I/E | 0.01 to 9.99 | +/- 1% | 0.01 |
| Inspiratory Average Pressure | FS: +/- 75 hPa | +/- 1% | 0.01 hPa |
| PEEP | +/- 75 hPa | 0.01 hPa | |
| Respiratory frequency | - | 0.01 cyde/min | |
| Ventilation | * | 10mL/min | |
| Index of Internal resistance | * | 0.01 hPa/L/s | |
| Index expiratory resistance to 0.5 L/s or to Demax/2 |
* | 0.01 hPa/L/s | |
| Resistance of the model | * | 0.01 hPa/L/s | |
| Elastance of the model | * | 0.01 hPa/L | |
| Compliance of the model | * | 0.01 mL/hPa | |
| Inspiratory time period | - | 0.01 sec | |
| Expiratory time period | - | 0.01 sec | |
| Work | * | 0.01 joule |
- : range of measure are not limited (temporary data)
* : range of measure limited by flow and pressure informations
The pneumotachograph 2 is proposed with the Rigel 355; others are available on request.
| Kind of pneumotachograph | Maximum Flow | Advised Flow |
| 00000 | 12 ml/s | 9 ml/s |
| 0000 | 20 ml/s | 15 ml/s |
| 000 | 60 ml/s | 40 ml/s |
| 00 | 150 ml/s | 100 ml/s |
| 0 | 350 ml/s | 250 ml/s |
| 1 | 1.2 l/s | 1 l/s |
| 2 | 3 l/s | 2.5 l/s |
| 3 | 8 l/s | 6.5 l/s |
| 4 | 14 l/s | 11 l/s |
| 5 | 25 l/s | 21 l/s |