• Other Seaward Sites:

Rigel 355 Ventilator Tester

Rigel 355 Ventilator Tester

The Most Complete Evaluation and Control Portable Tool

The Rigel 355 is a unique hand-held ventilator tester for both adult and paediatric ventilators. The Rigel 355 offers highly accurate measurements of: Respiration Frequency, Tidal volume, Maximum Pressure, Minimum Pressure, Ventilation, Expiratory average flow, Inspiratory average flow and Ratio I /E.

 

The Rigel 355 has a build-in dynamic memory to store performance graphs for download to your PC. When used with the included RESPI software, the Rigel 355 offers a full stand-alone ventilator test system which provides real time graphs on your PC.


The RESPI software provides additional parameters such as: Maximum Expiratory Flow, Tidal Volume of Inspiratory and Expiratory, Inspiratory Average Pressure, PEEP, Index of Internal and External Resistance, Resistance and Compliance of the Passive lung, Inspiratory and Expiratory Time.

Hand-Held

Paediatric and Adult
Ventilation

Battery Powered

Large LCD display

Customer Comments

No comments have been made about this product. Be the first to let us know what you think.

Knowledge Base

You must be registered to view Knowledge Base articles related to this product. Please register or log-in to view these articles.

Key Features

  • Respiratory Frequency
  • Ratio I/E
  • Tidal volume
  • Maximum Pressure
  • Minimum Pressure
  • Ventilation
  • Expiratory average flow 
  • Inspiratory average flow
  • Hi-Pressure Port (+/- 7Bar)
  • Low-Pressure Port (+/- 75 HPa)
  • Oxygen Level Measurement
  • Dynamic Memory
  • PC download

 

Downloads

Please feel free to download any of the documents associated with this product.

Technical Specification

Mecanical and electrical characteristics of Rigel 355
Dimensions : 225 x 115 x 80 mm ( + external pneumotachograph)
Weight : 0.900 Kg
Supply voltage : internal battery (12 hours autonomy)
mains supply with a 230 Volts adapter

Accessories
-Carrying suitcase
-Pneumotachographs (from n° 00000 to n° 5)
-Portable pr inter connected directly on Rigel 355
-Car charger
- Passive lung

Parameters directly calculated and displayed on Rigel 355 :

 Parameters  Symbol  measures

Precision

Display Resolution
 Respiratory Frequency  FR

 -

 -

 0.01 cycle/min
 Ratio I /E  I /E  0.01 to 9.99

 -

 0.01
 Tidal volume  Vt  Limited by the volume of the charge

 -

 10 mL
 Maximum Pressure  Pmax  FS : + /- 75 hPa    0.01 hPa
 Minimum Pressure  Pmin  FS : + /- 75 hPa  + /- 1%  0.01 hPa
 Ventilation  L/min  -  + /- 1%  10 mL/min
 Expiratory average flow  Fmoy  Depends on the used pneumo  + /- 1%  10 mL/s
 Inspiratory average flow  Fmoy  Depends on the used pneumo  + /- 1%  10 mL/s

Parameters displayed in Manometer / Flowmeter Mode:

 Parameters  Symbol  measures

Precision

Display Resolution
 Flow (pneumo type 2)  D

± 2,5 L/s or 150 L/min

2%  0.01 L/min
 Flow (pneumo type 0)  D  ± 0,25 L/s or 15 L/min 2%  0.1 mL/min
 Low Pressure  Pb  ± 60 hPa 1%  0.01 hPa
 High Pressure  Ph  0 à 5 bar s 1% 0.001 bar
 FiO2  O2

+ /- 100 %

1%

0.01 hPa

Calculated parameters by Rigel Respi software ( Rigel 355 used in Ventest Mode)

Parameters  Range of Measures   Precision Display Resolution 

Maximum Pressure
Minimum

 +/- 75 hPa  +/- 1%  0.01 hPa
Maximum expiratory flow
Inspiratory
 originally +/- 3 L/s (depends on
 the used pneumo, see table)
 +/- 5% (for 1 L volume)  10mL/s
 Expiratory Average flow
Inspiratory
 originally +/- 3 L/s (depends on
 the used pneumo, see table)
 +/- 5% (for 1 L volume)  10mL/s
Tidal Volume (Insp and Exp)  Depends on the volume of the charge    10mL
Ratio I/E  0.01 to 9.99   +/- 1%  0.01
 Inspiratory Average Pressure  FS: +/- 75 hPa   +/- 1%  0.01 hPa
 PEEP +/- 75 hPa    0.01 hPa
 Respiratory frequency -    0.01 cyde/min
 Ventilation  *    10mL/min
 Index of Internal resistance  *    0.01 hPa/L/s
 Index expiratory resistance to
 0.5 L/s or to Demax/2
 *    0.01 hPa/L/s
 Resistance of the model  *    0.01 hPa/L/s
 Elastance of the model  *    0.01 hPa/L
 Compliance of the model  *    0.01 mL/hPa
 Inspiratory time period  -    0.01 sec
 Expiratory time period  -    0.01 sec
 Work  *    0.01 joule

- : range of measure are not limited (temporary data)
* : range of measure limited by flow and pressure informations

 The pneumotachograph 2 is proposed with the Rigel 355; others are available on request.

Kind of pneumotachograph  Maximum Flow  Advised Flow 
 00000  12 ml/s  9 ml/s
 0000  20 ml/s  15 ml/s
 000  60 ml/s  40 ml/s
 00  150 ml/s  100 ml/s
 0  350 ml/s  250 ml/s
 1  1.2 l/s  1 l/s
 2  3 l/s  2.5 l/s
 3  8 l/s  6.5 l/s
 4  14 l/s  11 l/s
 5  25 l/s  21 l/s

Knowledge Base

Product Matrix

The Product Matrix allows you to compare the features of up to 5 products. Simply check the box next to the products you wish to compare then select the 'Compare Products' button.

Product Enquiry

Contact Information

indicates required field
Title*
Please type in the validation code from the image (case sensitive).

Additional Information (optional)

 
Site engineered by NorthBloc   
© 2010, Seaward Electronic Group Ltd. All rights reserved.