Rigel 322 SPO2 Simulator

Rigel 322 SPO2 Simulator

Pulse Oximetry Analyser

 

The Rigel 322 is a highly accurate, portable Pulse Oximetry Analyser using Patented Technology.

 

With its build-in Probe Analyser, small footprint, compact design and battery-powered operation, the Rigel 322 provides a unique balance between portability and easy-of-use.

 

The Rigel 322's patented use of Direct Electronic SPO2 simulation provides a highly accurate representation of the human's Oxygen Saturation level. With more than 13 preset R-curve available, the Rigel 322 proves to be a universal tool for those involved in testing the correct use of SPO2 monitors.


The large graphic display provides an instant and unambiguous picture of the continuity and quality of SPO2 probes.

 

 

Direct Electronic Simulation

Built-in Probe Analyser

Compact Design

 

Customer Comments

No comments have been made about this product. Be the first to let us know what you think.

Knowledge Base

You must be registered to view Knowledge Base articles related to this product. Please register or log-in to view these articles.

Key Features

 

  • Patented Electronic Simulation
  • 13 pre-set R-curves
  • Automatic and Manual Testing
  • User Selectable Patient Conditions
  • Probe Continuity Tests
  • Automatic Data Logging
  • Remote Control
  • Printing facilities
  • Large Graphic Display
  • Small and Light Weight
  • Battery and AC operation

Included with the Rigel 322:

  • Internal Rechargeable battery
  • Power Adaptor
  • Certificate of Calibration
  • Instruction Manual
  • Choice of 5 SPO2 Simulator / Probe cable sets*

*Simulator / Probe cable sets are available for all commonly used Pulse Oximeters. Please contact us with the type of Oximeters.

Downloads

Please feel free to download any of the documents associated with this product.

Technical Specification

Oxygen Saturation (SpO2)

Range - 0-100 SpO2%
Resolution - 1 SpO2%
Accuracy - ±0.5 SpO2%, 70-100 SpO2%
           ±1 SpO2%, 50-69 SpO2%

Pulse Rate

Range - 20-255 bpm
Resolution - 1 bpm
Accuracy - ±1 bpm, 20-199 bpm
           ±2 bpm, 200-255 bpm

R-Curves (Red/IR Ratio)

R-Curves specify the relationship between the Red/IR ratio measured by pulse oximeters and the SpO2 value displayed. R-curves are obtained by each pulse oximeter manufacturer during clinical trials. Preset Manufacturer's R-Curves (13): BCI, Criticare, CSI-Spot, Datascope, Datex, Hewlett-Packard, Masimo, Nellcor, Nihon Kohden, Nonin, Novametrix, Ohmeda & Sensormedics
Please contact the factory for the availability of other R-Curves.
 
Plethysmographic Amplitude (Simulates the effect of tissue perfusion)

Range - 0-200% of nominal amplitude
Resolution - 1%
Accuracy - ±1

Ambient Light (Simulates the ambient light level "seen" by the oximeter probe)
 
Range - DC Level & AC
Level - 0-255%
Resolution - 1%
Accuracy - ±1%
Flicker Frequency Presets (6) - 60 Hz
                          120 Hz
                          121 Hz
                          50 Hz
                          100 Hz
                          101 Hz

Motion Artifact

Amplitude Range - 0-100
Frequency - 4 Hz, Contact factory for availability of other frequencies.

Arrhythmias (Simulates cardiac arrhythmias by altering the plethysmographic pulse amplitude and pulse rate.)

Arrhythmias (5) - Asystolic Pause
            PVC
            Atrial Fibrillation
            Tachycardia
            Bradycardia

Auto Sequences

Preset Auto Sequences automate pulse oximeter testing.
Standard Presets (5) - CalDown - decreasing SpO2 and increasing Pulse Rate at 20 second intervals 
                                  CalUp - increasing SpO2 and decreasing Pulse Rate at 20 second intervals
                                  CalLong - decreasing SpO2 and increasing Pulse Rate at 30 second intervals 
                                  FullSwp - tests every SpO2 point between 60 and 100 at 21 second intervals
                                  Trend - Sleep Apnea pattern over 60 minutes.

Automatic Data-Logging & Printing

Closed-loop testing of pulse oximeters can be achieved by connecting a cable between the "SERIAL COMM" port of the oximeter and the "RS-232" port of the Rigel 322. Oximeter test results are automatically retrieved and displayed on the Rigel 322screen; the test results can be printed immediately or stored for future printing.
Serial data protocols supported: BCI, Criticare, Datascope, Datex, Hewlett-Packard, Nellcor, Nihon Kohden, Nonin, Novametrix & Ohmeda Please contact the factory for the availability of other serial data protocols.

Stored Printing

Stored Printing allows the user to save test results for printing at a later time or up-loading into popular equipment management programs such as Otis¨/Profile¨ and medTester/Sentinel.

Remote Control

The Rigel 322can be operated via remote control through its RS-232 port. In addition to duplicating all front panel operations, remote control commands include status query commands, a keypad lock command and a "warm-boot" command.

Probe Analyzer

Probe Continuity Test
Simultaneous Channels -  2 for Nellcor-style probes: LED & PhotoDiode circuits 
                                      3 for Ohmeda-style probes: Red LED, IR LED & PhotoDiode circuits
Test Mode - Constant Current Stimulus, Voltage Output
Constant Current Stimulus - 500 uA
Intermittent "Glitch" Detection - Detects open circuits and short circuits, 1 msec or longer
Probe Connection - Connectors for Nellcor & Ohmeda probes are built into Rigel 322; other probes are connected via adapter cables

General

Display
Type - Backlit graphic LCD
Resolution - 320 h x 240 v, 0.30mm dot pitch
Controls - Contrast, Backlight Timeout

Serial Interface
Type - RS-232C, 9 pin "PC-AT" style
Baud Rates - 300-9600 bps
Modes - Remote Control, Data-Logging & Equipment Management Up-Loading

Printer Interface
Type - Centronics, 25 pin "IBM-PC" style

Battery & AC Adapter

Battery
Type - 19.2V, Nickel Metal Hydride (NimH)
Capacity - 2.1A-Hr (About 4.5 Hours usage)
Charge Time - 17 Hours (charges even if Rigel 322in use)

External AC Adapter
Type - Wallmount Linear
Input - 120VAC, 60Hz or 230VAC, 50Hz
Output - 24VDC, 500mA
Safety Agency Approvals - UL, CSA, TUV

Physical

Case - Aluminum with non-skid feet & handle
Dimensions - 203 x 203 x 127 (WxDxH in mm)
Weight - <2 Kg including battery

Rigel 322 Cables

User's choice of 2 sets included in purchase price
Additional sets available at extra charge
SpO2 Simulator Adapter Cables - BCI, Criticare, CSI-Spot, Datascope, Datex, Hewlett-Packard, Masimo, Nellcor, Nihon Kohden, Nonin, Novametrix,Ohmeda & Sensormedics
Probe Analyzer Adapter Cables - Criticare, Datascope, Datex, Hewlett-Packard, Masimo, Nihon Kohden, Novametrix & Sensormedics
Note - BCI, Nellcor, Nonin & Ohmeda probes connect directly.

Optional Accessories

Automatic Data-Logging Cables - BCI, Criticare, Datascope, Datex, Hewlett-Packard, Nellcor, Nihon Kohden, Nonin, Novametrix & Ohmeda
Otis/medTester Interface - Rigel 322can up-load test results into Otis¨/Profile¨ and medTester/Sentinel equipment management systems.
Rigel 322 Carry Case

Knowledge Base

Product Matrix

The Product Matrix allows you to compare the features of up to 5 products. Simply check the box next to the products you wish to compare then select the 'Compare Products' button.

Product Enquiry

Contact Information

indicates required field
Title*
Please type in the validation code from the image (case sensitive).

Additional Information (optional)

 
Site engineered by NorthBloc