Rigel Medical offer a range of electrical safety analysers to provide you with improved efficiency and portability which can reduce the time spent at preventative or routine maintenance of medical devices. Our safety analysers are designed to meet the requirements of both National and local standards such as AAMI/ES 60601, UL2601, EN60601, IEC 62353, VDE0751, AS/NZ 3200, AZ/NZ 3551, MDA DB9801 - 2005 and many more.
The Rigel 288 is the FIRST universal truly hand-held medical electrical safety analyser to combine the features of an automatic / manual tester with a data logging / asset management facility. A compact lightweight design and long life battery power reduces downtime between tests, making the safety analyzer totally practical and highly portable for multi site use. In addition to IEC 60601-1 and AAMI / NFPA 99, the Rigel 288 safety analyser meets the newly published standard for in-service and after repair testing of medical electronic devices, the IEC 62353.
The Rigel 288’s large internal memory facilitates the storage of test results during preventative or routine maintenance, allowing easy download after tests are completed. Comprehensive database software is available to ensure fast and easy download of test results, managing your asset database, creation of test sequences, scheduling of Preventative Product Maintenance and producing test certificates. Wireless connection means that stored data can be transferred immediately and directly from the tester to PC-based record keeping systems at the touch of a button.
The Rigel 288 safety analyzer is aimed at those who require ease of use, portability and automation of electrical safety tests.
Rigel 288 comes as standard including:
Electrical Safety Tests performed
Specific to IEC 60601-1;
Specific to pr IEC 62353 – VDE 0751;
Specific to AAMI & NFPA;
Patient Leakage AP- ALL (AAMI / NFPA) Custom tests can be created using a variation or combination of the above.
* Using unique hi-current (>25A), low energy test current
RIGEL 288 SPECIFICATIONS
|
Earth Continuity |
|
| Method 2 wire technique | using 'zero' lead function. |
| Test Current | >+200mA -200mA DC into 2 ohms |
| Max Test Voltage | 4-24V rms o/c (6V for IEC 60601 tests) |
| Measuring Range (low range) | 0.001 – 0.999 ohms @ 0.001 ohms resolution |
| Measuring Range (mid range) | 1.00 – 9.99 ohms /@ 0.01 ohms resolution |
| Measuring Range (high range) | 10.0 – 19.9 ohms @ 0.1 ohms resolution |
| Accuracy | ± 3% of reading + 10 m ohms |
| Insulation Resistance | |
| Measurement | EUT to Earth / Ground |
| EUT to AP | AP to Ground |
| Voltage | 250V DC & 500 V DC @1mA. |
| Range (low range) | 0.01Mohms - 20 Mohms |
| Accuracy (low range) | ± 5% of reading +2 counts |
| Range (high range) @ 500VDC | 20Mohms – 100Mohms |
| Range (high range) @ 250VDC | 20Mohms – 50Mohms |
| Accuracy (high range) | ±10% +2 counts |
| Resolution | 0.01Mohms |
| Direct Leakage Measurement | |
| Measuring Range | 4μA to 9999μA |
| Accuracy | ± 5% or reading +2 counts |
| Mains on A.P. voltage | F-type only @ 110% of mains |
| Measuring Device | As per IEC 60601-1 requirements |
| Measurement Type | Separate AC & DC for Patient (-Auxiliary) Leakage to IEC 60601, True RMS for all remaining Leakage tests |
| Differential Leakage Measurement | |
| Measuring Range | 75μA to 9999μA |
| Accuracy | ±5% of reading + 5 counts |
| Measurement / display resolution | 1μA |
| Measurement Type | True RMS |
| Measuring Device | frequency response characteristics to IEC 60601-1. |
| Alternative Leakage Measurements: | |
| Test Voltage | 250V at mains frequency |
| Test Current | 3.5 mA current limited |
| Measurement Range | 4μA to 9999μA |
| Measurement Resolution | 1μA |
| Measurement Accuracy | ±5% of reading + 2 counts |
| Measurement Type | True RMS |
| Measuring Device | As per IEC 60601-1 |
| Power Measurement | |
| Method | VA rating. |
| Range | 0.1KVA – 4KVA |
| Accuracy | ±10% + 2 counts |
| Mains Outlet Test | |
| Input voltage range | 0-300V AC |
| max current | 16A |
| Measures | L – E, N – E |
| Accuracy | ± 5% of reading + 2 counts |
| IEC Mains Lead Test | |
| Test Duration | 2s |
| Test | Continuity of all EB conductor, Insulation & Polarity |
| General | |
| Mains power | 230 VAC ±10%, 50Hz +/- 1Hz |
| 120 VAC ±10%, 60Hz +/- 1Hz (USA model) | |
| Battery | 6 x 1.5V Alkaline AA |
| Weight | 1.6 kg including batteries |
| Size (L x W x D) | 270 x 110 x 75 mm / 10.5 x 4 x 3” |
| Operating conditions | 0˚C - 40˚C, 0-90% RH - NC |
| Storage environment | -15˚C - +60˚C |
| Environmental Protection | IP 40 |