.jpg)
Complete Electrical Medical Safety Analyser
The Rigel 277 Plus Electrical Safety Analyzer is a portable medical electrical safety analyser combining IEC/EN 60601-1 compliance with additional test facilities for IEC/EN 61010 (Laboratory Equipment) including Touch Leakage, Voltage Measurement and dedicated IEC 61010 Measuring Device (Body Model).
In addition, test routines are preset for MDA DB9801and VDE 0751. Customised test routines can be created to meet local standards and variants of safety test standards such as AS/NZ 3200 and previous editions of IEC / EN 60601 standards.
The Rigel 277 Plus Safety Analyser performs manual, automatic and semi automatic test routines allowing full control of power up and power down delay times. The combination of a full graphics display, large internal memory and integral qwerty keyboard provides a complete stand alone test system to proof the electrical safety of both medical and non medical electrical equipment.
The Rigel 277 Plus Safety Analyzer is designed to increase the efficiency of test routines and to reduce the overall test time
Included with the Rigel 277 Plus:
Applications
|
AC Earth Bond Test |
|
| Test Current | 100mA, 1A, 10A, 25A |
| Current Source | Regulated |
| Open Test Voltage | 6V rms nominal |
| Range | 0.01ohms - 19.99ohms / 0.01ohms counts |
| Accuracy | ± 5% of reading, ± 2 counts |
| Measurement | 4-wire, floating earth |
|
|
|
| Insulation Test | |
| Test Voltage | 350 / 500V d.c. nominal |
| Auto Ranging | 100kohms - 20 Mohms (L) / 0.01Mohms counts |
| 20Mohms – 100Mohms (H) / 0.01Mohms counts | |
| Accuracy | ± 5% of reading, ± 2 counts (L) |
| ±10% of reading, ± 2 counts (H) | |
| Open Circuit Voltage | <750VDC |
|
|
|
| IEC 60601 Leakage Test | |
| Supply Voltage | Incoming mains (230/110VAC) |
| F-Type Open Circuit Voltage | 110% +/-20% of Mains Input Voltage |
| Maximum Load | 16A (13A for UK model) |
| Input Impedance | 1Kohm – as per 60601 MD |
| Frequency Response | as per IEC 60601-1 requirements |
| Range | 0.000mA – 9.999mA / 0.001 counts |
| Accuracy. | ± 3% of reading, ± 4 counts |
|
|
|
| IEC 61010 Leakage Test | |
| Supply Voltage | Incoming mains (230/110VAC) |
| Maximum Load | 16A (13A for UK model) |
| Input Impedance | 2Kohms – as per 61010 MD |
| Frequency Response | DC – 2.5kHz (as per IEC 61010) |
| Range | 0.10mA – 3.50 mA / 0.01 counts |
| Accuracy. | ± 3% of reading, ± 4 counts |
|
|
|
| Load Test | |
| Measured Load | 0.06 - 4kVA |
| Measured Voltage Mains Supply | +/-10% |
|
|
|
| IEC Lead Test | |
| Test | 40V a.c., 1mA nominal |
| Detects | Good, Open, Short, Reverse |
|
|
|
| General | |
| Size | 400mm x 320mm x 160 mm |
| Weight | appr. 9 kg |
| Operating | 0ºC to 40ºC (non condensing) |
| Storage | -10ºC to 50ºC (non condensing) |
| Maximum | R.H 90% |