• Other Seaward Sites:

Rigel 266 Plus Electricall Safety Analyser

Rigel 266 Plus Electricall Safety Analyser

Easy to Use Medical Safety Analyser

 

The Rigel 266 Plus is a highly compact, easy-to-use medical electrical safety analyser, designed to test in accordance with IEC/EN 60601-1, MDA DB9801, and AS/NZ 3200.

In addition to the Semi Automatic test, the Rigel 266 Plus safety analyzer provides fully manual test routines, easily selectable via the rotary selector switch and dedicated push buttons for individual Single Fault Conditions. A sequence of leakage measurements can be selected without requiring a power break, providing a quick test solution for medical equipment requiring a power-up / down delay time.

A large LCD display provides clear indication of test results, Single Fault Condition, AC or DC measurements and pass/fail limits.

With its compact design and full IEC 60601-1 features, the Rigel 266 Plus safety analyser provides the user with a highly effective and portable test solution.

 

Included with the Rigel 266 Plus:

  • Supplied with rugged carry case
  • RS 232 for real time download into PC / Laptop*
  • Earth Bond Probe
  • Applied parts harness including Applied Parts Adaptors

* PC / Laptop must be isloated from earth

 

 

Customer Comments

No comments have been made about this product. Be the first to let us know what you think.

Knowledge Base

You must be registered to view Knowledge Base articles related to this product. Please register or log-in to view these articles.

Key Features

  • Portable & Lightweight 
  • Full IEC 60601 Leakage Tests
  • Up to 25A AC Test Current 
  • Large Graphic LCD 
  • Separate AC & DC Values 
  • Full Manual Control 
  • Insulation Test 
  • 5 Patient Connections 
  • Dedicated IEC Lead Test 
  • Connects to optional printer

  Applications 

In-service Safety Testing of: 

  • Patient Monitors, Infusion Pumps, Hospitals Beds, X-ray and Ultrasound Equipment, Video Equipment, PC based Medical Systems, Defibrillators and many other Medical and non-Medical Electrical Equipment and Systems.

  • Fast & Efficient Testing of IEC leads
 
  • Earth Bond Testing on (Medical) Installations and non-Medical Equipment Portable Appliance Testing
 

Downloads

Please feel free to download any of the documents associated with this product.

Technical Specification

AC Earth Bond Test

 
   
Open Test Voltage 6V rms nominal
Test Current 1A, 10A, 25A AC
Range 40mohms – 19.99ohms
Resolution. 0.01ohms
Accuracy 5) { alert("You may only select 5 products to compare at any one time."); return false; } return true; }

Product Matrix

The Product Matrix allows you to compare the features of up to 5 products. Simply check the box next to the products you wish to compare then select the 'Compare Products' button.

Product Enquiry

Contact Information

indicates required field
Title*
Please type in the validation code from the image (case sensitive).

Additional Information (optional)

 
Site engineered by NorthBloc   
© 2009, Seaward Electronic Group Ltd. All rights reserved.