Easy to Use Medical Safety Analyser
The Rigel 266 Plus is a highly compact, easy-to-use medical electrical safety analyser, designed to test in accordance with IEC/EN 60601-1, MDA DB9801, and AS/NZ 3200.
In addition to the Semi Automatic test, the Rigel 266 Plus safety analyzer provides fully manual test routines, easily selectable via the rotary selector switch and dedicated push buttons for individual Single Fault Conditions. A sequence of leakage measurements can be selected without requiring a power break, providing a quick test solution for medical equipment requiring a power-up / down delay time.
A large LCD display provides clear indication of test results, Single Fault Condition, AC or DC measurements and pass/fail limits.
With its compact design and full IEC 60601-1 features, the Rigel 266 Plus safety analyser provides the user with a highly effective and portable test solution.
Included with the Rigel 266 Plus:
* PC / Laptop must be isloated from earth
Applications
In-service Safety Testing of:
|
AC Earth Bond Test |
|
| Open Test Voltage | 6V rms nominal |
| Test Current | 1A, 10A, 25A AC |
| Range | 40mohms – 19.99ohms |
| Resolution. | 0.01ohms |
| Accuracy | +/- 5% of reading, +/- 2 counts |
| Measurement | 4-wire measurement, floating earth |
|
Insulation Test |
|
| Test Voltage | 500V d.c. nominal (0.5Mohms load) |
| Range | 100kohms - 100 Mohms |
| Resolution | 0.01 Mohms (<100 Mohms) |
| Accuracy. | +/- 5% of reading, +/- 2 counts |
| (100kohms - 20Mohms) | |
| All Leakage Tests | |
| Input Impedance | >1Mohm |
| Frequency Response | as per IEC 60601-1 requirements |
| Accuracy | +/- 5% of reading, +/- 10 counts |
| Range | |
| Earth Leakage | 0.04mA – 9.999mA |
| Enclosure Leakage | 0.04mA – 9.999mA |
| Patient Leakage | 0.004mA – 9.999mA |
| Patient Auxiliary Leakage | 0.004mA – 9.999mA |
| Patient SIP / SOP Leakage | 0.04mA – 9.999mA |
| Patient F-Type Leakage | 0.025mA – 9.999mA |
| F-Type Open Circuit Voltage | 110% +/-20% of Mains Input Voltage |
| Load Test | |
| Measured Load | 0.06 - 4kVA |
| Measured Voltage | Mains Supply +/-10% |
| IEC Lead Test | |
| Test | 40V a.c., 1mA nominal |
| Detects | Good, Open, Short, Reverse |
| Mechanical | |
| Size | 310mm x 210mm x 80mm |
| Weight | 4kg |
| Environmental | |
| Operating | 0ºC to 40ºC (non condensing) |
| Storage | -10ºC to 50ºC (non condensing) |
| Maximum R.H | 90% |