Rigel 277 Plus

Universal Automatic Safety Analyser

  • IEC / EN 60601-1 / 61010
  • On-board printer and memory
  • 0.1 to 25A Earthbond current
  • Up to 25A AC Test Current
  • Full IEC 60601 Leakage Tests
  • Full IEC 61010 Leakage Tests
  • Multi Class Applied Part
  • Large Graphic LCD
Rigel 277 Plus
Click to Request a Demonstration / Quote

Rigel 277 Plus

The Rigel 277 Plus is a portable medical electrical safety analyser combining IEC/EN 60601-1 compliance with additional test facilities for IEC/EN 61010 (Laboratory Equipment) including Touch Leakage, Voltage Measurement and dedicated IEC 61010 Measuring Device (Body Model).

Customised test routines can be created to meet local standards and variants of safety test standards. The Rigel 277 Plus safety analyser performs manual, automatic and semi automatic test routines allowing full control of power up and power down sequence, thus reducing the overall test times.

The Rigel 277 Plus boasts a large full graphics display, internal memory, QWERTY keyboard and printer making this a completely stand alone medical safety analyser.

Rigel 277 Plus

Technical Specification:

AC Earth Bond Test
Test Current 100mA, 1A, 10A, 25A
Current Source Regulated
Open Test Voltage 6V rms nominal
Range 0.01ohms - 19.99ohms / 0.01ohms counts
Accuracy ± 5% of reading, ± 2 counts
Measurement 4-wire, floating earth
Insulation Test
Test Voltage 350 / 500V d.c. nominal
Auto Ranging 100kohms - 20 Mohms (L) / 0.01Mohms counts
20Mohms – 100Mohms (H) / 0.01Mohms counts
Accuracy ± 5% of reading, ± 2 counts (L)
±10% of reading, ± 2 counts (H)
Open Circuit Voltage <750VDC
IEC 60601 Leakage Test
Supply Voltage Incoming mains (230/110VAC)
F-Type Open Circuit Voltage 110% +/-20% of Mains Input Voltage
Maximum Load 16A (13A for UK model)
Input Impedance 1Kohm – as per 60601 MD
Frequency Response as per IEC 60601-1 requirements
Range 0.000mA – 9.999mA / 0.001 counts
Accuracy ± 3% of reading, ± 4 counts
IEC 61010 Leakage Test
Supply Voltage Incoming mains (230/110VAC)
Maximum Load 16A (13A for UK model)
Input Impedance 2Kohms – as per 61010 MD
Frequency Response DC – 2.5kHz (as per IEC 61010)
Range 0.10mA – 3.50 mA / 0.01 counts
Accuracy ± 3% of reading, ± 4 counts
Load Test
Measured Load 0.06 - 4kVA
Measured Voltage Mains Supply +/-10%
IEC Lead Test
Test 40V a.c., 1mA nominal
Detects Good, Open, Short, Reverse
General Specification
Size 400mm x 320mm x 160 mm
Weight Appropriately 9 kg
Operating 0ºC to 40ºC (non condensing)
Storage -10ºC to 50ºC (non condensing)
Maximum R.H 90%
Part Numbers
350A910 UK / Ireland
350A913 Euro - Schuko
350A917 French
350A915 Danish
350A918 Czech / Polish
350A914 US (110V – 60Hz)
350A912 Australian

Barcode Scanner