Rigel 266 Plus
Compact Manual Electrical Safety Analyser
- IEC / EN 60601-1
- Compact and robust enclosure
- 1-25A Earthbond current
- Portable & Lightweight
- Full IEC 60601 Leakage Test
- Up to 25A AC Test Current
- Large LCD Display
- Separate AC & DC Value
![]()
![]()
![]()
![]()
![]()
![]()
The Rigel 266 Plus is the most compact electrical safety analyser capable of delivering 25 amps earthbond current and testing in accordance with IEC/EN 60601-1, MDA DB9801, and AS/NZ 3200.
The Rigel 266 Plus safety analyser boasts a large LCD screen and rotary selector, and test routines can be manually controlled for quick and easy testing of medical devices. Single Fault Conditions can be manually introduced to minimise power breaks and reduce test times.
With its compact design and full IEC 60601-1 features, the Rigel 266 Plus safety analyser provides the user with a highly effective and portable test solution.
Rigel 266 Plus
| AC Earth Bond Test | |
|---|---|
| Open Test Voltage | 6V rms nominal |
| Test Current | 1A, 10A, 25A AC |
| Range | 40mohms – 19.99ohms |
| Resolution | 0.01ohms |
| Accuracy | +/- 5% of reading, +/- 2 counts |
| Measurement | 4-wire measurement, floating earth |
| Insulation Test | |
|---|---|
| Test Voltage | 500V d.c. nominal (0.5Mohms load) |
| Range | 100kohms - 100 Mohms |
| Resolution | 0.01 Mohms (<100 Mohms) |
| Accuracy | +/- 5% of reading, +/- 2 counts (100kohms - 20Mohms) |
| All Leakage Tests | |
|---|---|
| Input Impedance | >1Mohm |
| Frequency Response | as per IEC 60601-1 requirements |
| Accuracy | +/- 5% of reading, +/- 10 counts |
| Range | |
|---|---|
| Earth Leakage | 0.04mA – 9.999mA |
| Enclosure Leakage | 0.04mA – 9.999mA |
| Patient Leakage | 0.004mA – 9.999mA |
| Patient Auxiliary Leakage | 0.004mA – 9.999mA |
| Patient SIP / SOP Leakage | 0.04mA – 9.999mA |
| Patient F-Type Leakage | 0.025mA – 9.999mA |
| F-Type Open Circuit Voltage | 110% +/-20% of Mains Input Voltage |
| Load Test | |
|---|---|
| Measured Load | 0.06 - 4kVA |
| Measured Voltage | Mains Supply +/-10% |
| IEC Lead Test | |
|---|---|
| Test | 40V a.c., 1mA nominal |
| Detects | Good, Open, Short, Reverse |
| Mechanical | |
|---|---|
| Size | 310mm x 210mm x 80mm |
| Weight | 4kg |
| Environmental | |
|---|---|
| Operating | 0ºC to 40ºC (non condensing) |
| Storage | -10ºC to 50ºC (non condensing) |
| Maximum R.H | 90% |
| Part Numbers | |
|---|---|
| 350A910 | UK (13A) |
| 350A913 | Schuko (16A) |
| 350A915 | Danish (10A) |
| 350A917 | French (16A) |
| 350A918 | Czech / Polish (16A) |
Rigel’s Med-eCare Plan is at your service for the entire Rigel Medical range. It covers all typical and unexpected costs associated with calibration and maintenance of your test equipment for the duration of the plan. View Product
The battery powered results printer is one of the most compact printers on the market. View Product
The Rigel 601 checkbox accurately replicates the leakage characteristics of equipment designed to IEC 60601-1, providing calibrated leakage values for the verification on electrical safety analysers. View Product