Rigel 601 Calibration Checkbox
- Calibrated leakage and resistance values
- 36 month recalibration interval
- Cost effective solution
![]()
The Rigel 601 checkbox accurately replicates the leakage characteristics of equipment designed to IEC 60601-1, providing calibrated leakage values for the verification on electrical safety analysers. With the ability to generate both AC and DC leakage currents as well offering dedicated F-type circuitry, the Rigel 601 provides calibrated values for Earth Leakage, Enclosure Leakage, Patient Leakage and Patient Leakage Mains on Applied Parts.
In addition, the Rigel 601 is able to provide a series of highly accurate resistance values to check the linearity of the Earth Bond and Insulation measuring circuits.
Dedicated switch positions provide individual PASS and FAIL values for B/BF and CF limits as per IEC 60601-1.
Rigel 601 Calibration Checkbox
| Earth Bond | |
|---|---|
| Test Points | 0.05 Ohm 0.10 Ohm 0.20 Ohm |
| Accuracy | +/- 2%, nominal value, factory calibrated |
| Maxumum Test Current | 25 A |
| Maximum Test Duration | 30 seconds @ 25A |
| Insulation | |
|---|---|
| Test Points | 2M Ohm 20M Ohm |
| Accuracy | 1%, nominal value, factory calibrated |
| Leakage | |
|---|---|
| Earth Leakage | Yes |
| Enclosure Leakage | Yes |
| Patient Leakage | Yes |
| PASS Value | 85% of IEC 60601 test limit |
| FAIL Value | 115% of IEC 60601 test limit See Table T1 for values |
| Accuracy | 0.5%, factory calibrated |
| F-type Leakage | |
|---|---|
| PASS Value | 25μA |
| FAIL Value | 2.5mA |
| Accuracy | Nominal value |
| General Specification | |
|---|---|
| Weight | 3.2KG |
| Size | 250 x 140 x 200mm |
| Calibration Interval | 36 months |
| Warranty | 12 months |
| Operating Conditions | |
|---|---|
| Temperature | 15-30°C |
| Humidity | 0-98% RH Non condensating |
The Rigel 266 Plus is the most compact electrical safety analyser capable of delivering 25 amps earthbond current and testing in accordance with IEC/EN 60601-1, MDA DB9801, and AS/NZ 3200. View Product
The Rigel 277 Plus is a portable medical electrical safety analyser combining IEC/EN 60601-1 compliance with additional test facilities for IEC/EN 61010 (Laboratory Equipment). View Product
The Rigel 288 has become the most recognised electrical safety analyser on the market today. View Product
The Rigel 62353 is the latest hand-held safety analyser from Rigel Medical, able to test to the new standard, the IEC 62353 (2007). View Product